A sensor for the backlight permanently determines the luminance of the monitor. The benefit: The defined and calibrated values are rendered exactly just seconds after the monitor is turned on and remain constant during the entire period of use. The sensor is invisibly integrated in the monitor.
The MX216 has a factory preset DICOM characteristic and displays gray tones of object-oriented radiological images. The monitor displays a typical dark room contrast of up to 1500:1 and therefore allows for clear and high-contrast image rendering.
Thanks to the high 2 Megapixels (colour) resolution, a strong contrast ratio of 1500:1 and stable brightness of up to 500 cd/m2, the monitor offers excellent image quality. Even the differences between the finest details are shown – regardless of your viewing angle. This is a great advantage if multiple physicians are looking at the screen.
The MX216 is the first MX-Series monitor to include EIZO's Hybrid Gamma PXL function, which automatically distinguishes between monochrome and colour images pixel by pixel, creating a hybrid display where each pixel has optimum grayscale. As a result, monochrome images are displayed in the DICOM® Part 14 grayscale, while colour images are reproduced corresponding to Gamma 2.2, for improved efficiency when viewing colour and monochrome images together.
LCD panels with a high brightness level tend to have more blurry image rendering thanks to over-framing than would be possible in comparison with an acquired exposure. Therefore, EIZO offers blur reduction anchored in monitor hardware. It retrieves details lost in the contours on the screen, meaning that the image is rendered as clearly as possible.
Colour rendering is controlled by a 13-bit look-up table (LUT), up to 10 bits of which are available in the DisplayPort connection. This produces a resolution with a maximum of 1 billion colour tones. The rendering characteristic and fine structures required for diagnostics can therefore be precisely identified.
The monitor shines thanks to its high colour purity and uniform illumination. This is down to the Digital Uniformity Equalizer (DUE), which corrects imbalances automatically, pixel by pixel. Gray and colour tones of radiological and other medical images are correctly rendered over the entire display. This is vital for diagnostics.
The precise calibration of white point and tone value characteristic curve is provided by an integrated front sensor (IFS). This measures the brightness and grayscales and calibrates the monitor autonomously according to the DICOM standard. The sensor works automatically, without restricting the field of vision of the monitor. You can save the costs, time, and effort of maintenance and rely on a consistently balanced image quality.
The display properties, in particular brightness and contrast, are suited to the creation of image rendering systems compliant with DIN 6868-157. The DICOM® GSDF characteristic is already precisely configured in the factory. This means that greyscales are consistent, which is vital for diagnostics.
Radiologists face an increasing amount of information on their screens, as modalities become increasingly digitalized. But imaging procedures now no longer need to be so complex, thanks to the unique Work-and-Flow technology from EIZO, featuring new functions developed specifically to meet the requirements of radiologists. The RadiForce MX216-SB and RadiCS-LE software solution enable you to benefit from the Work-and-Flow functions.
The Point-and-Focus function allows you to quickly select and focus on relevant image areas using your mouse or keyboard. By adjusting the brightness and greyscale, you can highlight interesting parts of an image by dimming the surrounding areas.
Manage Effortless Calibration
|
With the RadiCS LE software (included) and Integrated Front Sensor (IFS) built into the front bezel, you can easily calibrate to DICOM Part 14 without having to mount, run, and remove an external sensor.
Simple calibration using the monitor backlight sensor is also supported.
The MX216-SB can be operated in both portrait and landscape format. The ‘Image Rotation Plus’ function, included in the RadiCS LE software provided, rotates the displayed image automatically depending on the position of the monitor. (This function only works if supported by the graphics board.)
In addition, the flexible stand guarantees optimal ergonomics. You can tilt the monitor or lower it down to desk level.

Thanks to the Image Rotation Plus function, the displayed image rotates automatically into portrait or landscape format, depending on the position of the monitor.
It doesn't get easier than this: You can connect most of your devices, such as PC, laptop or cameras directly to the monitor because the monitor has a number of different ports. That makes your daily work easier.
The monitor has an automatic shut down option for the backlight (backlight saver). This extends the duration of use. Similar to a screen saver, the LEDs turn off when the screen is not being used.
Thanks to the signal input and output, you can link several RadiForce monitors through their DisplayPort interface. This means that you can realise multi-monitor solutions with the greatest of ease – without laborious and excessive cabling.
For environments, in which clinical images are reviewed with applications for electronic patient files, the MX216 provides a cost-efficient solution, which supports the DICOM standard and enables significant savings.
EIZO measures and adjusts each tone of gray carefully so that the monitors comply with the DICOM standard when delivered from the factory. The result is a particularly consistent gradation of gray tones, allowing for optimal radiological clinical reviews.
Space-Saving Design for Clinical Environment
|
Compared to the previous MX215, the monitor's width, height, and depth are reduced by 3.4 mm, 22.7 mm, and 45.5 mm respectively – a total of 23% less space – saving more workspace for other tasks in clinics and other facilities.
- 2 megapixel LCD module with LED backlight for reliably high and permanently stable brightness
- Pallet with 68 billion color tones for precise color rendering up to 12 bit
- Automatic control of the luminance distribution (Digital Uniformity Equalizer)
- Prepares for calibration, acceptance and constancy testing in accordance with DIN 6868-157, and quality assurance guidelines
- Integrated sensor for automatic and exact calibration
- Presence sensor for immediate readiness when there is someone in front of the screen
Model Variations | MX216-BK: with stand, black |
---|---|
Panel | |
Type | Color (IPS) |
Backlight | LED |
Size | 54 cm / 21.3" |
Native Resolution | 1200 x 1600 (3:4 aspect ratio) |
Viewable Image Size (H x V) | 324.0 x 432.0 mm |
Pixel Pitch | 0.270 x 0.270 mm |
Display Colors | 10-bit colors (DisplayPort): 1.07 billion (maximum) colors 8-bit colors: 16.77 million from a palette of 543 billion colors |
Viewing Angles (H / V, typical) | 178° / 178° |
Brightness (typical) | 500 cd/m2 |
Contrast Ratio (typical) | 1500:1 |
Response Time (typical) | 20 ms (on / off) |
Video Signals | |
Input Terminals | DisplayPort, DVI-D |
Output Terminals | DisplayPort (daisy chain) |
Digital Scanning Frequency (H / V) | 31 - 100 kHz / 59 - 61 Hz |
USB | |
Upstream | USB 2.0: Type-B |
Downstream | USB 2.0: Type-A x 2 |
Power | |
Power Requirements | AC 100 - 240 V: 50 / 60 Hz |
Typical Power Consumption | 26 W |
Maximum Power Consumption | 55 W |
Power Save Mode | 0.6 W or less |
Sensor | Backlight Sensor, Integrated Front Sensor, Ambient Light Sensor |
Features & Functions | |
Brightness Stabilization | Yes |
Digital Uniformity Equalizer | Yes |
Hybrid Gamma PXL | Yes |
Preset Modes | CAL Switch |
OSD Languages | English, German, French, Italian, Japanese, Simplified Chinese, Spanish, Swedish, Traditional Chinese |
Physical Specifications | |
Net Weight | 7.6 kg |
Net Weight (Without Stand) | 4.7 kg |
Hole Spacing (VESA Standard) | 100 x 100 mm |
Certifications & Standards (Please contact EIZO for the latest information.) | CE (Medical Device), EN60601-1, ANSI/AAMI ES60601-1, CSA C22.2 No. 601-1, IEC60601-1, VCCI-B, FCC-B, CAN ICES-3 (B), RCM, RoHS, China RoHS, WEEE, CCC, EAC |
FDA | 510(k) Clearance for General Radiography |
Dedicated Software | |
Monitor Quality Control Software RadiCS | Supported |
Supplied Accessories (May vary by country. Please contact EIZO for details.) | |
Signal Cables | DisplayPort (3 m) |
Others | AC power cord (3 m), USB cable (3 m), Utility Disk (RadiCS LE, PDF instructions for use, PDF installation manual), instructions for use |
Warranty | Five Years |
Dimension Drawing |